| Bulletin
          2/2005 ---------- ............ Stellenanzeigen Tagungskalender Impressum Vorstand SGSMP(Adressen)
 ---------- andere Bulletins / d'autres bulletins / altri bolletini /
        other bulletins
         | 
          
            |  ...
        are named after the German physicist Georg Christoph Lichtenberg.
        One possibility to create such figures is to irradiate acrylic with a
        beam of high energy electrons. As the charge builds within the insulator, the effective voltage can reach millions of
              volts. Finally the electrical stress exceeds the dielectric strength of the
              plastic, causing it to suddenly become conductive in a process called dielectric
              breakdown. Acrylic is easy to get hold of - but a high energy electron beam?
        This is when  being a medical physicist is definitely an advantage
        - at least if one is working in radiation therapy: we all use high
        energy electrons from a linear accelerator to treat patients, either
        directly, or to produce x-rays for treatment. If you are in this lucky
        situation, you should not rush to repeat our experiments, however: the
        dose rate available in clinical mode is insufficient to build up a
        sufficient charge within a short enough time; not all linear
        accelerators on the market are suitable for this exercise - and the
        suitable ones need to be modified. You must not try to repeat our
        experiments with a linac due to treat patients the next day - unless you
        have a large number of expensive spare parts sitting on your shelves.
  Why? |  
          
            |   |  |  
            | Look at the effect of the light field mirror
              (left) and the monitor
        chamber (right) to know why. The effect on the electronics was less easy
        to document - but similarly impressive! The next time you decommission
        your Varian Clinac 2100 (I cannot comment on other manufacturers' linacs),
              however, this is your chance. |   UNDER NO CIRCUMSTANCES SHOULD YOU REPEAT THE 
        PROCEDURE DESCRIBED BELOW ON A
        LINEAR ACCELERATOR WHICH IS STILL IN CLINICAL USE !In the electron mode at maximum rep rate irradiation times of
            approximately 20 min are required to build up the necessary charge
            in the acrylic. In photon mode (6 MV) with the target removed,
            however, only some 6 to 10 seconds were needed during our experiment: 
          As the acrylic block will later be placed on the treatment couch,
            rotate the gantry to lateral (90 or 270 degree, depending on the
            camera position for watching the event).The linac needs to be tuned in this exact position. As the monitor
            chamber will not deliver correct signals under these circumstances,
            and it will be damaged soon after the start of the irradiation anyway (see above picture) all
            steering servos
            are switched off (PFN and AFC servos stay on). It is therefore very important that the linac
            remains stable with the steering servos switched off.Now (with the beam switched off) the target is pulled out of the beam path, and fixed.Also the flattening filter is removed (as it would otherwise act as
            target).The acrylic is placed on the table on some wood (for additional
            insulation). At this stage the cross hair can still be used: mark
            the position of the block, and possibly note different table
            positions for blocks of differing sizes, as the light field mirror
            will also be destroyed pretty soon after the start of the
            irradiation; the lateral laser would also be
            useful for positioning - but if you want to keep using it, it might
            be a good idea to remove it first.Zoom in the TV camera on the acrylic, and switch off the lights
            before switching the beam on.One has to work in the service mode, to override the dosimetry and
            target interlocks.CAUTION: Due to the extreme dose rate not only the monitor
            chamber, but also the vacuum window might be damaged if the beam is
            switched on after these modifications!Watch the acrylic glow when the beam is switched on.The result is difficult to predict - find out by trial and error
            what you like best. Parameters are the block dimensions, holes
            drilled into the block, and beam-on time (a few comments are added
            below).In our experience it is not necessary (well ... not possible) to
            invoke the discharge using an earthed pin or similar object: even if the
            beam was switched off before the discharge occured "automatically",
            the discharge had occured by the time we were in the room.CAUTION: Even after discharge there will be residual
            charges on or within the acrylic! Before touching the block with
            your fingers, use an earthed cable on a long stick to remove at
            least the major part of this residual charge ... and don't be
            surprised to experience a prickly sensation nevertheless. 
         
         
          
            |   |   |  
            | The result of a long evening (the Clinac stood up to this extreme
        torture much longer than expected) ... you are invited to have a closer look
        at some of the samples: click on images to see them with higher
              resolution. |  
          Disclaimer: The fact that it worked
        for us is no guarantee that it will work somewhere else under different
        circumstances. Whatever you do, including taking any precautions
        considered necessary, it is entirely your responsibility.
            |  | Block 10 x 10 x 3 cm3 6 MeV electrons have a range of about 1.5 cm in Perspex; a 3
              cm thick block is therefore ideal to get a Lichtenberg figure
              roughly in the middle of the block (irradiate perpendicular to the
              10 x 10 cm2 face); a small hole (1.5 mmdia
              and about 1 mm deep) was drilled in the centre of a small side to provoke
              a discharge to this point. The hole not only defines the position,
              it also results in a discharge at lower charge, creating less
              damage to the acrylic, and producing a finer structure of the
              Lichtenberg figure.
 |  
            |  | Block 5 x 13 x 3 cm3 Hole drilled in the centre of the large side (white dot on the
              photograph); irradiated from the opposite side.
 |  
            |  | Block 7 x 7 x 5 cm3 Irradiation from two sides. First irradiation from the
              "bottom" (on the photograph) with no hole drilled. The
              result of the first irradiation looked similar to the first
              photograph. with an arbitrary focus point sideways (= bright spot
              on the photograph). For the second irradiation the block was
              rotated, with the discharge being attracted by the damage in the
              acrylic caused by the first discharge.
 This is the original photograph used for the "title picture"
              above.
 |  
            |  | Block 5 x 5 x 5 cm3 Irradiation as above; viewed from two sides. Whilst the
              first irradiation (from "bottom") discharges to a side,
              the second irradiation (from "top") always discharges to
              the existing paths created during the first irradiation.
 |  |  
            |  | Block 5 x 5 x 5 cm3 For the second irradiation (from the "top" of the
              photograph) irradiation was continued after the first spark,
              resulting in many fine lines due to almost continuous discharges
              between the paths of the original discharge. Not visible on the
              photograph is a strong yellow tint on the beam entry side of the
              second exposure, identical to what is usually seen on long used
              acrylic shielding block trays in radiotherapy.
 |  
            |  | Block 7 x 7 x 5 cm3 First irradiation (from "bottom" of photograph)
              with a comparably high dose (for some reason the discharge
              occured later than usual) resulting in some severe damage (lower
              right corner of photograph); the second irradiation resulted in
              two simultaneous discharges along different paths.
 |  
            |  | Block 5 x 13 x 3 cm3 Deeper hole (about 5 mm) drilled on one side; despite this a
              second simultaneous discharge to a second focus point on the
              opposite side occured.
 |  
            |  | Block 5 x 13 x 3 cm3 Same block as above, viewed from the side.
 |  |